What to do if Titan Logix Corp Measuring Instruments input voltage is too high?
M
Michael KirbySep 23, 2025
If the Titan Logix Corp Measuring Instruments indicates that the input voltage is too high, check the power supply for adequate operating voltage.
K
Kelly WagnerAug 31, 2025
Why is there a faint beeping coming from the Titan Logix Corp TD100 horn?
C
Christopher RodriguezAug 31, 2025
If you notice a faint beeping coming from the horn of your Titan Logix Corp Measuring Instruments, check the horn for loose, corroded, or disconnected wiring. Also, ensure that a 150ohm to 500ohm, 2W resistor is connected across the horn.
R
Robert RussellAug 25, 2025
Why does my Titan Logix Corp Measuring Instruments display restart at erratic intervals?
M
madelineterryAug 25, 2025
If your Titan Logix Corp Measuring Instruments display restarts at erratic intervals, check the power supply wiring for intermittent or corroded connections. Also, ensure that the battery voltage is at least 8VDC and steady. If the issue persists, the Finch II Display may be faulty, and you may need to replace the unit.
H
hurleymatthewAug 25, 2025
What causes loss of power in Titan Logix Corp TD100?
H
hlesterAug 26, 2025
If you experience a loss of power and no system operation with a blank display on your Titan Logix Corp Measuring Instruments, check the fuses, cable connections, wiring, and power switches. Ensure the power supply provides adequate operating voltage. Use a DMM to test points for battery voltage. If fuses keep blowing, verify that the fuse type and rating are appropriate for the application.
M
Mark MillerAug 26, 2025
How to fix Titan Logix Corp Measuring Instruments display that does not turn on?
K
kimsophiaAug 26, 2025
If your Titan Logix Corp Measuring Instruments display does not turn on, check for a blown power fuse and inspect the power or ground wiring for loose, corroded, or disconnected wires.
If your Titan Logix Corp Measuring Instruments display shows erratic level measurements or '2 LO', physically separate all TD100 system components and wiring from radio devices. Check the probe for physical damage, defects, corrosion, or product build-up. The TD100 transmitter may be defective. Also, check the probe to TD100 transmitter connection for contamination.
B
Bianca JuarezAug 27, 2025
What causes incorrect volume display on Titan Logix Corp TD100?
M
Monique VelasquezAug 27, 2025
If the Titan Logix Corp Measuring Instruments shows incorrect volume, offset calibration may be required. Also, the TD100 transmitter may be incorrectly programmed or defective.
D
Danny SkinnerAug 28, 2025
What causes TD80 or TD100 timeout in Titan Logix Corp Measuring Instruments?
J
John HawkinsAug 28, 2025
If you are experiencing a TD80 or TD100 timeout with Titan Logix Corp Measuring Instruments, check the transmitter wiring on the appropriate compartment. The TD series transmitter or the display itself might be defective, so checking the wiring is crucial.
C
Christopher EllisAug 29, 2025
Why do the fuses keep blowing on Titan Logix Corp TD100?
M
michaelrasmussenAug 29, 2025
If the Titan Logix Corp Measuring Instruments fuses keep blowing, verify that the type and rating of the fuse is appropriate for the application. Also, verify that the system components are installed and wired correctly.
L
Laura JacobsAug 29, 2025
What does ‘- - - -’ mean on Titan Logix Corp Measuring Instruments display?
S
Shannon ArellanoAug 29, 2025
For a Titan Logix Corp Measuring Instruments displaying only ‘- - - -‘, check the wiring between the display and transmitter.