EasyManua.ls Logo

TOOLCRAFT 1713198 - Calibration and Maintenance; Calibrating the Microscope

TOOLCRAFT 1713198
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
14. Calibrating the Microscope
Focus the microscope on the calibration ruler.
Rotatethenefocusdialuntilthepicturebecomessharp.
Adjust the microscope position so the image is lined
up square with the XY guides.
Double-click an image in the saved images window (B7) to enter the measurement interface, then select
the calibration icon.
Click and drag to select start and end points across
a known value e.g. 1 mm increment along the ruler.
When the dialog box shows, enter the exact value of
the measurement sample in the Actual dimension
eld.Thesoftwarewillcalibratethemagnication.
Select OK to conrm. The adjusted magnication
value will show in the upper right corner.
15. High and Low Magnication
Highmagnication(1.5mmeldofview):Turnthenefocusdial(A6) all the way to the right, then slowly dial back
until the image comes into focus.
Lowmagnication(6mmeldofview):Turnthenefocusdial(A6) all the way to the left, then slowly dial back until
the image comes into focus.
Afterturningthenefocusdial(A6) all the way right or left, it will take a few seconds for the LED light
balance to adjust.