ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES for RF emissions class A
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The IS-5500 is intended for use in the electromagnetic environment specified below. The customer or the user of the IS-
5500 should assure that it is used in such an environment.
Emissions Tests Compliance Electromagnetic Environment Guidance
The IS-5500 uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
The IS-5500 is suitable for use in all establishments other than domestic
and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations
IEC 61000-3-3
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The IS-5500is intended for use in the electromagnetic environment specified below. The customer or the user
of the IS-5500 should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic
Environment Guidance
(ESD)
IEC 61000-4-2
±8kV air
±8kV air
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
Transient/Burst
IEC 61000-4-4
lines
±1kV for input/output
supply lines
±1kV for input/output
Mains power quality should
be that of a typical
commercial or hospital
IEC 61000-4-5
±2kV common mode
mode
±2kV common mode
Mains power quality should
be that of a typical
commercial or hospital
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
T
U
T
) for 0,5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
T
U
T
) for 0,5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the IS-5500requires
continued operation during
power mains interruptions, it
is recommended that the IS-
5500 be powered from an
uninterrupted power supply