2031005680 Rev. C, 06/2016 ii
If possible, perform cleaning/disinfection of parts between usage by different patients using
alcohol swab.
There are no periodic maintenance requirements for the IS-5500. However, periodic checks
of all the functions and safety switches should be performed every six months.
WARNING:
SAFETY CONCERNS
The IS-5500 Stand conforms to IEC 60601-1 3rd Ed, AAMI ES 60601-1 and CSA C22.2 No.
60601-1 certifications. Any of the optional equipment defined in this manual can be added to the IS-
5500 without introducing any additional safety hazards.
There is no contraindication to use of this stand.
The stand does not have applied parts. However, for the parts that may contact the patients, the
requirements of Type B applied parts are applied.
WARNING: The installation of the IS-5500 stand shall only be performed by a
qualified technician.
WARNING: The IS-5500 stand is designed for usage with TOPCON equipment.
If any other equipment is used, it shall meet the same certifications.
WARNING: All service (except replacing the two external fuses) and repairs
shall be performed by a qualified technician.
WARNING: All replacement parts or optional equipment shall be genuine
Topcon parts. Contact Topcon or your dealer for those items.
!
!
!
!
!
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective
earth
Connecting electrical equipment to MSO effectively leads to creating an ME SYSTEM, and can result in a
reduced level of safety.
No modification of this equipment is allowed.
An additional MULTIPLE SOCKET-OUTLET or extension cord cannot to be connected to STAND.
Connect to the stand only items that have been specified as part of ME SYSTEM or specified as being
compatible with ME SYSTEM
MULTIPLE SOCKET-OUTLETS provided with the ME SYSTEM to be used only for supplying power to
equipment intended to form part of ME SYSTEM
RISKS OF HIGH LEAKAGE CURRENT, connect only equipment specified to be connected through the
stand and forming part of the ME System
DO NOT touch pins of hand held instrument charging wells and patient simultaneously.
To safely switch off the stand, press the control panel power switch to off, and disconnect power plug from
outlet.
In the case of liquid entering into the stand, allow time to dry, and perform thorough inspection of device.
Do not lean or push onto the stand during usage.