Topcon RL-VH4DR/G2 - Horizontal Calibration Check

Topcon RL-VH4DR/G2
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
31
5
Turn the unit on again and allow auto-leveling to
complete.
6
Make a new mark (Mb) where the laser beam strikes
the paper.
7
Measure the distance between the first mark (Ma)
and the second mark (Mb). No calibration is
necessary if distance is within 6 mm (1/4 of an inch).
8
Repeat procedure for the Y axis.
If less than 6 mm (.23 inches)
No calibration Necessary
30 m (98 feet)
X2 laser beam
X1 laser beam