33
3
When the beam is precisely centered, press the [ENTER] key.
The [X] LED will flash.
When the flashing stops, the X axis calibration adjustment is made and power is turned off.
4
For Y axis calibration, turn the unit as instructed in step 1 then press the X/Y Axis Selection
key.
Confirm that the [Y] LED is lit, then repeat steps 2 to 4 for the Y axis.
Repeat the checking procedure to confirm proper calibration has been made.
This operation can be performed with the Remote Controller RC-40.
When using the RC-40, press the [ENTER] key on the RC-40 unit.
(The [ENTER] key for the RC-40: The same mark as the [ENTER] key for the
instrument.)
This operation can be performed with the Remote Controller RC-40.
When using the RC-40, select the Y axis using the X/Y axis selection key and move the
laser beam up and down with the Alignment control key.
NOTE
If the calibration is greater than the adjustment allows, the error LED will start flashing.
(See “Error Code” on page 51)
If this occurs, contact your Topcon dealer.
When the laser beam is difficult to see, it can be adjusted in the Laser sensor mode or
the Laser pointing mode.