(2) Adjusting Calibration
Turn the unit off by pressing the [START] key once. Confirm that unit has shut off before
beginning the following procedure. (In step 2 and 3, optional RC-40 remote controller is helpful.)
1
Without moving the unit, press the [ENTER] and [START] keys simultaneously.
2
Press either the right or left key on the Alignment Control key to move the laser beam up or
down on wall B until the measurement for the distance dHb is the same as the measurement
dHa on wall A.
This operation can be performed with the Remote Controller RC-40.
When using the RC-40, select the Y axis using the X/Y axis selection key and
move the laser beam up and down with the Alignment control key.
START
1
XV
Y
2
3
ENTER
ERROR
4
X/Y
Laser beam
up or down