3.3 pH Calibration Messages
1. If a 1-point calibration has been successfully performed, the
Aqua-pH will display the following message, and then display the
asymmetry and slope of the sensor. If the meter has not been
calibrated at two points at this stage, the slope is set to 100.0%.
1 point 6.88
Cal. OK
then:
Asym= 0.10pH
Slope=100.0%
2. If a 1-point calibration has failed, the Aqua-pH will display the
following message, then the failed asymmetry value of the sensor.
1 point 6.88
Cal. Failed
then:
Asym= 1.50pH
3. If a 2-point calibration has been successfully performed, the
Aqua-pH will display the following message, and then the asymmetry
and slope of the sensor.
2 point 4.00
Cal. OK
then:
Asym= 0.10pH
Slope= 99.5%
4. If a 2-point calibration has failed, the Aqua-pH will display the
following message, and then the failed slope value of the sensor.
2 point 4.00
Cal. Failed
then:
Slope= 70.0%