Operating Guide | TS-700 Series
15
PARAMETER
FUNCTION
OPTIONS
C15
Inner Code
display
Enter C15 to check the
inner code.
C16
Date
Set the date, from left to
right: year/month/day
C17
Time
Set the time from left to
right: hour/min./sec.
C18
Communication
setting
(Serial interface
data output
method)
0 = Serial interface data
output disabled
1 = Continuous sending,
remote display
2 = Print key sends data to
printer or PC.
3 = PC command request
(DataLogPC software)
4 = PC continuous
sending.
5 = Remote display
continuous sending
format.
C19
Baud rate
0 = 1200 1 = 2400
2 = 4800 3 = 9600
C20
CAUTION:
Changing
these
parameters will
affect the
indicator’s
performance.
Zero range
(amount that
can be zeroed
out with the
Zero key)
0 = Disable manual zero
setting
1 = ±1% max capacity
2 = ±2% max capacity
4 = ±4% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
100 = ±100% max capacity
C21
Zero range at
power-up
0 = no initial zero setting
1 = ±1% max capacity
2 = ±2% max capacity
5 = ±5% max capacity
10 = ±10% max capacity
20 = ±20% max capacity