Operating Guide | TS-700 Series
A-12
Appendix For Scale Technicians
Data Output Formats
REMOTE DISPLAY FORMAT
State A
Bits: 0,1,2
0
1
2
Decimal point
position
1
0
0
XXXXXX0
0
1
0
XXXXXXX
1
1
0
XXXXX-X
0
0
1
XXXX-XX
1
0
1
XXX-XXX
Bits: 3,4
Division
0
1
X1
1
0
X2
State B
BITS
FUNCTION
Bits0
Gross = 0, net=1
Bits1
Symbol: positive = 0,negative =1
Bits2
Overload (or under zero)=1
Bits3
Dynamic = 1
Bits4
Unit: lb=0, kg=1
Bits5
Constant 1
Bits6
Constant 0
State C
Bit2
Bit1
Bit0
unit
0
0
0
Kg or lb
0
0
1
g
0
1
0
t
Bit 3
printing=1
Bit 4
Extend
display=1
Bit 5
Constant 1
Bit 6
Constant 0