A small calibration switch is located at the bottom
left corner of the circuit board. Move the switch to
the “On” position. Close the enclosure.
ALL TS-700 MODELS: After enabling access to
calibration parameters, proceed as follows:
CALIBRATION PROCEDURE
1) Power on the indicator.
2) Press the PRINT and HOLD keys. “C” will
display. Release the keys.
3) “C01” will display, with the 1 flashing. The
indicator is now is Parameter Settings mode.
Navigating in Settings Mode
When the indicator is in settings mode, refer to the
graphic printed below the key to determine the keypad
key’s function.
The and arrow keys are used to
increase/decrease the value of the flashing digit, or to
navigate up and down a list of preset options.
The and arrow keys are used to move the active
digit over to the left or to the right.
The key and key can be used to move
forward or backward through the parameters. Also, the
key is used to enter/accept changes after they
have been made to the parameter settings.
The key can be used to exit settings mode.
With the indicator in settings mode and the cali-bration
weight nearby, proceed with the calibration process
while referring to the following chart: