XD ONE USER MANUAL44
BreatheLITE is designed to return a “Pass”/“Fail” result based on the compliance audit check
results and operational threshold for the applicable particulate sizes. This is displayed on
completion of the check and is detailed in the device log of each connected XD One.
Return of a “Pass” result
On return of a “Pass” result, the particulate sensor is functioning as expected and normal
monitoring operation can resume. Return of a “Fail” result.
Return of a “Fail” result
1. Run a sensor cleaning operation as highlighted in section 15.2.
2. Repeat the compliance audit check sequence and note the test result.
3. On return of a “Pass” result, the particulate sensor is functioning as expected and normal
monitoring operation can resume.
4. On One unit return of a repeat “Fail” result, please contact Trolex direct to discuss support or
servicing of the XD One.
15.6 Compliance audit check: Results
In some circumstances, a routine preventative maintenance schedule should be used to ensure
that the performance of the device is upheld. The following table should be used as guidance to
the level of unit maintenance required based on environmental dust loadings.
Note: All particulate and operating environment types are dierent and therefore this matrix should
be used as a guide only. It is recommended that an assessment of the site environmental and
operating conditions is carried out periodically to determine the required frequency of a routine
maintenance schedule.
15.7 Preventative maintenance
Dust loading Average dust loading Expected maintenance
in mg/m³ schedule
Medium Up to 10 mg/m³ 3 to 6 months
Low Up to 5 mg/m³ 6 to 12 months
High 10 mg/m³ or above 1 to 3 months
It is recommended that the XD One is operated with location and proximity consideration relating to
atomising dust suppression systems. Instrument readings will include atomised or misted sizes that
pass through the XD One sensor, within the particle detection range.
15.8 Atomised particulate suppression and mist spray
XD One_User manual_P5644.1601_Rev B