EN Operating manual – Material moisture measuring device T660 6
Configuration mode
1. Press the Left/menu key (11) for approx. 2 seconds.
– The device emits a short beep.
– The CFG symbol (12) is displayed in the upper right corner.
– The available options for configuration mode (17) are dis-
played.
2. Use the cross control (2) to select the desired option.
3. Confirm the selection by pressing the OK button (10).
– The selected symbol will be illuminated.
Setting the alarm
Here you can determine the limit value for the alarm function.
Upon exceeding this value, the device emits an acoustic signal
and the indication ALARM (17) flashes. The alarm function draws
on the current digit measured value. The limit value can be deter-
mined within a range of 0 to 200.
1. Select ALARM (17) whilst in configuration mode.
2. Press the OK button (10) to confirm.
– The upper measurement value display (14) flashes.
3. Press the Up (7) or Down (9) key to activate or deactivate the
alarm.
– On or Off will be indicated in the upper measurement value
display (14).
4. Press the Right/back key (8).
– Depending on the selection, the alarm is either activated or
deactivated.
– The lower measurement value display (16) flashes.
5. Press the keys Left/menu (11) or Right/back (8) to select a
digit.
– The selected digit flashes.
6. Press the Up (7) or Down (9) key to change the value of the
selected digit.
7. Repeat the steps 5. and 6. until the value is set as desired.
8. Press the OK button (10) for approx. 2 seconds.
– The alarm was set based on your choice.
– The device switches to measuring mode.
– With activated alarm function the indication ALARM (17)
continues to be displayed.
Adjusting the display illumination
The display illumination can be adjusted within a range of 20 to
100 %. Another available setting is AL.on. AL.on features a
brightness of 100% and deactivates the automatic switch-off
function.
1. Select the lamp (17) whilst in configuration mode.
2. Press the OK button (10) to confirm.
3. Select the desired value by use of the Up (7) or Down (9) key.
4. Press the OK button (10) for approx. 2 seconds.
– The set value will be adopted.
– The device switches to measuring mode.
Configuration mode Description
ALARM Setting the threshold value for the alarm
Lamp Adjusting the display illumination
MAT Material settings
CAL Setting the offset