3-6
T
T
a
a
k
k
i
i
n
n
g
g
A
A
s
s
p
p
h
h
a
a
l
l
t
t
M
M
e
e
a
a
s
s
u
u
r
r
e
e
m
m
e
e
n
n
t
t
s
s
1. Place the gauge on the surface of the asphalt, choosing a
smooth surface for best results. Sand can be used to fill large
voids. Care must be taken to ensure that the gauge is resting
on the asphalt and not on the sand. This procedure is only
helpful when the voids are large.
2. Place each hand on opposite corners of the base to check for
rocking.
3. Turn the depth indicator switch to BS.
4. Turn the PWR/TIME switch to the FAST (15 second)
position or the NORMAL position.
5. Place the source rod in the backscatter position, making
sure the source rod is in the BS notch and not resting on the
asphalt.
6. Press MEASURE. The window will show the gauge
counting and ERR in the upper-left corner.
7. After the counting stops, press WD and read the Wet
Density.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler 3400-B Series and is the answer not in the manual?

Troxler 3400-B Series Specifications

General IconGeneral
BrandTroxler
Model3400-B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals