Model 3400-B Series
Appendix A-9
3411-B Error Codes
Error
Code
Indicated Failure
Mode
Probable Cause
01
System failed internal
test condition.
Microprocessor memory board
hardware failure.
02
Accumulated number
exceeds display size.
Position of PWR/TIME switch
was changed during
accumulation OR hardware
failure on I/O board.
10,
11
Bad keyboard input.
Two keys depressed, defective
keyboard or hardware failure
on I/O board.
30
Microprocessor tried to
evaluate the log of a
negative number.
Incorrect depth selected by
DEPTH switch.
30
Microprocessor
attempted to divide by
zero.
No standard counts in the MS
or DS registers.
31
Negative overflow in
division.
Incorrect depth selected by
DEPTH switch OR standard or
measure counts are not valid.
32
Positive overflow in
division.
Same as above.
40
Invalid input from
MOISTURE
CORRECTION switches.
Switch failure, I/O board
failure, or operator induced
error during chain calculations.
41
Invalid input from
DEPTH switch.
Same as above.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler 3400-B Series and is the answer not in the manual?

Troxler 3400-B Series Specifications

General IconGeneral
BrandTroxler
Model3400-B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals