42
be used in the integral. If an end-trigger is set (set with
SETx
±nnn.nn), the acquisition will stop either when the end-trigger
condition is detected or when nnnn samples has been acquired,
whichever comes first. After the command is finished, a termination
sequence is sent to signal the end of the transfer.
Example 4)
DAFxx0005 Request 5 samples of flow in comma
delimited ASCII format (assume begin-trigger is not
set).
Flowmeter returns the acknowledgement “OK” <CR>
<LF> followed by the flow data and termination
sequence 1.10,1.20,1.25,1.23,1.20,1.10 <CR> <LF>
Example 5) DCFTxx0003 Request 3 samples of flow and
temperature in comma / <CR> delimited ASCII format
(assume begin trigger is not set).
Flowmeter returns the acknowledgement “OK” <CR>
<LF> followed by (flow) comma (temperature) <CR>
<LF> (flow) comma (temperature) <CR> <LF> (flow)
comma (temperature) <CR> <LF>
After importation into a spreadsheet it might look like
this: