What to do if TSI Instruments Measuring Instruments analog output is not in proportion to display?
- RRichard CollinsAug 25, 2025
If the analog output of your TSI Instruments Measuring Instruments is not in proportion to the display, the analog output range in the DUSTTRAK™ monitor may be set incorrectly, or the data logger scaling factor may be set incorrectly. Check the analog output setting in the Setup->Analog screen, ensuring the channel of interest is selected and the correct output (0 to 5V, 4 to 20 mA) is selected. Review the scaling factor set in the Setup-Analog screen.