DEFAULT command. This command also clears both the begin- and
end-trigger values.
Example: The normal factory default parameter settings are:
Analog Pressure Setting: 101.3 kPa
Sample Rate: 10 ms
Gas Calibration: 0 = Air
Flow Units: Standard
Analog Output Scaling: The full scale flow rate of the
meter.
Analog Zero Intercept: 0 mV
Beginning Trigger Disabled
End Trigger Disabled
Depending on the configuration of your flowmeter, these default
settings may be different.
After the command is processed, an acknowledge sequence of “OK”
<CR> <LF> is sent. If the command generated an error, an error code
of “ERRn” <CR> <LF> will be sent. See the end of this section for a
list of possible error codes.