EasyManua.ls Logo

TSI Instruments 9565 Series - Page 49

TSI Instruments 9565 Series
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications 43
Power Requirements:
Four AA-size batteries (included) or
AC Adapter p/n 801761
Input: 90 to 240 VAC, 50 to 60 Hz
Output: 9 VDC, 2A
1
Temperature compensated over an air temperature range of 40 to 150°F (5 to
65°C).
2
The accuracy statement of ±3.0% of reading or ±3 ft/min (±0.015 m/s), whichever is
greater, begins at 30 ft/min through 9999 ft/min (0.15 m/s through 50 m/s).
3
Pressure velocity measurements are not recommended below 1000 ft/min (5 m/s)
and are best suited to velocities over 2000 ft/min (10.00 m/s). Range can vary
depending on barometric pressure.
4
Accuracy is a function of converting pressure to velocity. Conversion accuracy
improves when actual pressure values increase.
5
Accuracy with instrument case at 77°F (25°C), add uncertainty of 0.05°F/°F
(0.03°C/°C) for change in instrument temperature.
6
Accuracy with probe at 77°F (25°C). Add uncertainty of 0.1% RH/°F (0.2% RH/°C)
for change in probe temperature. Includes 1% hysteresis.
7
Overpressure range = 190 in. H
2
O (7 psi , 360 mmHg, 48 kPa).

Table of Contents

Related product manuals