>> Ultrasonic Flow meter User Manual
29
4) Clean the surface of installation points
Paint, rust and anti-corrosive coating on installation points need to be cleaned. It’s good to use a
polishing machine to get the metal luster.
As shown below:
5) Install transducers
After transducer wiring and sealing, please evenly smear 2-3mm couplant on the transducer
emitting surface. Then put the transducers on the installation points, fixed with steel belt or steel
rope.
6) Check Installation
Please see details in Chapter 7.5