9
5.
Zero setting
(In case of PR-32
α
, PR-101
α
, PR-201
α
)
CAUTION
The refractometer needs the "zero setting" (calibration) before it is used for the first
time in a day. Moreover, if the ambient temperature changes during the daily work with
the refractometer, it needs the zero setting again.
The zero setting is also needed whenever the battery is replaced.
Do not use any metallic implement for dropping water on the prism because the metal
tool possibly damages the prism surface.
3.5
0 0
0
3.5
[ 1
]
Flashes 3 times
Measurement
value is d
isplayed
Zero-set correctly.
Completely cover prism surface
Min. 0.1mL
Clean the prism
surface thoroughly.
Apply tap water.
Press the
START/OFF Button.
Measurement:not 0.0%
Measurement:0.0%
Press the
ZERO Setting
Button.
Zero setting is complete.
Flashes
3 times
000 is
displayed.
Perform zero-setting
Confirm zero-setting
20.4
0
0
.
0
20.4
0
0
.
0
20.4
0
0
.3
3.5
0 0
0