The preset reminder alarm is ringing Press any button to turn o the alarm or wait for the alarm to
be turned o automatically in 30 seconds.
Temperature is TOO HIGH during the test procedure. Large variation may occur between results due to high or
low temperature. Move to an environment
(50°F ~ 104°F or 10°C ~ 40°C) and wait for 30 minutes before
re-testing.
Battery power is LOW. Meter will only provide
proximately 10 more measurements.
Replace with 2 AAA batteries.
1. An error message indicates that the hematocrit of
the blood sample is too low for HT100-B system to
perform a test.
2.Use control solution to check the meter and test
strips.
1.Follow the instructions and try again with a new test strip.
Please contact customer service if the problem persists.
2.Follow the instructions to enter control solution mode, and
try again with a new test strip.
Temperature is TOO LOW during the test procedure. Large variation may occur between results due to high or
low temperature. Move to an environment (50°F ~ 104°F or
10°C ~ 40°C) and wait for 30 minutes before re-testing.
Display Description
Action
An error message indicates that the hematocrit of
the hematocrit of the blood sample is too high for
HT100-B system to perform a test.
Follow the instructions and try again with a new test strip.
Please contact customer service if the problem persists.
29