Display Messages
The system is ready to accept a blood sample. Apply a blood sample.
The result is HIGHER than the “HI” alert setting. The default setting is 180 mg/dL and it can be changed
according to HI/LO Alarms Setup
The default setting is 70 mg/dL and it can be changed
according to HI/LO Alarms Setup
The result is LOWER than the “LO” alert setting.
Display Description Action
Test result is HIGHER than 600 mg/dL. Repeat the test by using a new test strip. If the same
message displays again, the result is conrmed to be higher
than the meter test range. Please consult your healthcare
professional immediately for further advice.
Test result is LOWER than 20 mg/dL.
Repeat the test by using a new test strip. If the
same message displays again, the result is
conrmed to be lower than the meter test range.
Please consult your healthcare professional
immediately for further advice.
28