CALIBRATION INSTRUCTIONS
NOTE: Scale can only be accurately calibrated using standard calibrated scale weights. Using
non-calibrated weights may result in inaccurate scale calibration. Nothing should be on the scale and
scale should be level.
ACTIVATE WEIGHT CALIBRATION MODE
Switch the mini jumper SWA1 to "ADJ" position. Adjust is down; Lock is up. The switch is located under
a panel on the right side of the keypad housing.
Turn on the scale. As scale counts down to zero, press and hold the ZERO key until the software
version (ex: 02018 ) appears on the display.
Release ZERO key. F 0 appears in display.
Press ENTER key. Display shows ZEro.
Press ENTER key. Display shows 0---,-Ck g.
Press UNITS or Kg key to set scale to lb or kg as needed.
Use the UP and ENTER keys to preset the weight being used to calibrate the scale. The UP key sets the
number and the ENTER key moves right one digit.
Example: To set a weight of 3.0 kg, you would press the ENTER key three times to move right three
spaces and the UP key three times to enter the number 3.
Place your calibrated weight on the scale.
Press ENTER key. Display shows WAIt.
Scale displays weight and beeps.
Scale is calibrated, it will now return to weighing mode automatically.
Switch the mini jumper SWA1 back to "LOCK."
0
0
Kg
UNITS
1
-800-295-551
0
uline.com

Other manuals for U-Line H-4593

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the U-Line H-4593 and is the answer not in the manual?

U-Line H-4593 Specifications

General IconGeneral
BrandU-Line
ModelH-4593
CategoryScales
LanguageEnglish

Related product manuals