Never apply voltage to the inputs.
Temperature affects readings. Allow the CLM100 to attain the ambient temperature of the wire under test.
Use a 9 Volt alkaline battery only
Use the test leads that came with the unit only. Other leads will not work.
Ensure that the test leads are clean and in good working order.
Use an abrasive pad similar to the one provided to ensure the wire under test is clean and free of oxidation
Ensure that insulation is stripped adequately from both ends of the wire under test. This will enable
measurements to be accurate.
Ensure that the alligator clips are connected across the diameter of the wire under test.
When measuring wire for insertion into conduit, add extra wire to compensate for the accuracy of the
meter. For example, 205 feet of wire is needed to run inside a piece of conduit that is 200 feet long. (This
would allow 2.5 feet on each end to attach the wire.) At 205 feet and 70˚F, the cable length meter has an
accuracy of ±7 feet. In this case it would be safer to measure out 212 feet of wire. This would ensure that
enough wire is available for the application.