No,SK00-7397-EI-003-06
PIRANI VACUUM GAUGE
MODEL GP-1G
MODEL GP-1G WITH CASE
(Pa/Torr)
INSTRUCTION MANUAL
This manual is for the following gauges.
Serial Nos. 02000G and higher
Read this manual before operation and keep it
at your hand for immediate reference.
Components Division,
ULVAC , Inc.
http://www.ulvac.co.jp/

Questions and Answers

  • A
    Ann RogersAug 4, 2025
    What to do if the Ulvac GP-1G meter needle is deflected off-scale?
    • W
      William ScottAug 4, 2025
      If the meter needle remains deflected off-scale to the right on your Ulvac Measuring Instruments, it could be due to several reasons: * The pressure might be beyond the measurable range (0.4 to 2700 Pa). * The sensor head cable may not be connected; reconnect it after turning the power off. * The sensor head cable might be disconnected; check its continuity with a circuit tester and replace it if necessary. * The filament might have burnt out; check its continuity with a circuit tester (using no more than 10 mA current to avoid shortening its life; resistance should be about 13 ? at atmospheric pressure) and replace it if needed.
  • R
    Richard SalazarAug 10, 2025
    Why does the needle vibrate on Ulvac GP-1G Measuring Instruments and not indicate a fixed value?
    • J
      jasonhunterAug 10, 2025
      If the needle vibrates and doesn't indicate a fixed value on your Ulvac Measuring Instruments, consider these possible causes: * The sensor head type may not be the specified one; try changing it or recalibrating the current one. * The sensor head might be contaminated, or its filament may be worn out; consider replacing the sensor head. * There could be a leak in the sensor head or the installation area; replace the sensor head if it's leaking, or stop the leak elsewhere. * Increased wire resistance due to poor contact or corrosion of the sensor head cable; check and replace the cable if needed. * The line voltage may not be within 100 to 240 VAC; ensure the voltage is within this range.
  • M
    Melody JonesAug 17, 2025
    What to do if pressure indication on Ulvac Measuring Instruments does not exceed 2 kPa at atmospheric pressure?
    • J
      Jasmine CaldwellAug 17, 2025
      If the pressure indication on your Ulvac Measuring Instruments does not exceed 2 kPa when measuring atmospheric pressure, consider the following: * The sensor head type may differ from the specified one; replace it with the correct one or adjust/recalibrate the current one. * The gas under measurement may not be nitrogen, as the gauge is calibrated for nitrogen. * The sensor head may be contaminated, or its filament may be worn out; replace the sensor head. * The sensor head or controller might be installed in a high-temperature location (the gauge was calibrated at approximately 25°C); move it to a clean, well-ventilated area without drafts.
  • K
    Kelly BakerAug 22, 2025
    How to fix Ulvac GP-1G when the meter needle doesn't deflect on power ON?
    • M
      Martin Calhoun DDSAug 22, 2025
      If the meter needle of your Ulvac Measuring Instruments does not deflect when the power is turned on, the fuse may have burnt out. Replace the fuse (250 VAC, 0.5 A) after turning off the power to the controller. If the fuse blows out immediately after replacement, the device requires repair.
  • B
    Brittney HudsonAug 28, 2025
    What to do if Ulvac Measuring Instruments pointer doesn't move initially but starts moving after pressure decreases?
    • L
      Leah JacksonAug 28, 2025
      If the pointer on your Ulvac Measuring Instruments doesn't move from the atmospheric pressure position initially but starts moving suddenly after the pressure decreases, it may be due to the gauge being exposed to high temperature and high humidity for an extended period. Ensure the gauge is used under the conditions listed in the specifications. If the problem persists in a regular environment, the gauge needs repair.

Summary

Safety Cautions

OVERVIEW

SPECIFICATIONS

Details of the vacuum gauge's technical parameters and capabilities.

DESCRIPTION OF COMPONENTS

Identification and explanation of the GP-1G gauge's physical components.

GP-1 G

Front view of the GP-1G gauge, detailing meter and zero adjustment.

GP-1 G with case

Front view of the GP-1G gauge with case, showing controls and indicators.

INSTALLING THE PIRANI GAUGE

INSTALLATION

Procedures for physically installing the vacuum gauge.

OPERATING PROCEDURE

TROUBLESHOOTING

APPENDIX

ACCURACY OF PIRANI VACUUM GAUGE

Details on the gauge's accuracy and error range.

WARRANTY