What to do if the Ulvac GP-1G meter needle is deflected off-scale?
- WWilliam ScottAug 4, 2025
If the meter needle remains deflected off-scale to the right on your Ulvac Measuring Instruments, it could be due to several reasons: * The pressure might be beyond the measurable range (0.4 to 2700 Pa). * The sensor head cable may not be connected; reconnect it after turning the power off. * The sensor head cable might be disconnected; check its continuity with a circuit tester and replace it if necessary. * The filament might have burnt out; check its continuity with a circuit tester (using no more than 10 mA current to avoid shortening its life; resistance should be about 13 ? at atmospheric pressure) and replace it if needed.