APPENDIX
- 21 -
5.3. Accuracy of Pirani Vacuum Gauge
The relationship between pressure and output of a Pirani vacuum gauge is nonlinear and its accuracy
range varies with pressure values. To express the outgoing inspection criterion for a Pirani vacuum
gauge, it is defined as "within ± 3% of 100% full scale as converted to the linear scale " and is given
in the instruction manual as measurement accuracy.
To express the error range by pressure values
The pressure and recorder output of the GP-1G(RY) is in proportion to the controller meter scale
graduations, as shown by the recorder output characteristic table (graph).
What is meant by "within ± 3% of 100% full scale as converted to the linear scale " is that an error of
± 3% of the linear scale is involved when the meter S range scale is replaced by a linear scale of
100% full scale.
To express the error range at a reference pressure of 40 Pa (0.4Torr) by pressure value, it is 32 to 49
Pa (0.33 to 0.48Torr), which is ± 3% as converted to a linear scale, as shown in Fig. 8.
Fig. 8 Linear memory conversion
The "Error Range of Pirani Vacuum Gauge" table is given for reference.
Indicated Values (Pa)
Working Standard(Pa) Indicated Values (Pa)
(Torr
Indicated Values (Torr)