22
Model UT50D: OPERATING MANUAL
Measurement Operation(10)
l For testing the capacitor with polarity, connect the
red test lead or test clip to anode & black test lead
or test clip to cathode
l When the tested capacitor is shorted or the capacitor
value is overloaded, the LCD display “1”.
l To minimize the measurement error caused by the
distributed capacitor, the testing lead or testing clip
should be as short as possible.
l To ensure precision especially when measuring small
capacitance 200nF range, the correct reading is as
follows:
Measurement displayed reading minus the Meter’s
open circuit displayed value.
l It is normal to take a while for zeroing when changing
over the measurement range. This process will not
affect the accuracy of the final readings obtained.