EasyManua.ls Logo

UNI-T UT522

UNI-T UT522
21 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UT522 USERS MANUAL
16
6.2 Simple measurement(use simple 2-Wires test leads for measurement):
This method will be used when the auxiliary earth stake cannot be used. Use
the expose earth resistance object from the ground as earth, like metal water
bath, water pile, wire ground, building earth, also can use 2-wires test leads
method (E and P&C terminals).Connection as figure 3:
figure 3
Primary
Side
Red
Green
Secondary
Side
Supply Transformer
Peference Earthed
Electrode
Earthed Electrode
Under
Test

Related product manuals