12
PressENTER button to confirm.The instrument will go from previous wavelength
(546nm) to the desired wavelength (500nm) and automatically blank.
Note: You must blank your reference before measure any sample. Follow your lab
procedures for preparing the reference liquid and the steps below to set the blank:
2.2.4 Make a blank reference solution by filling a clean cuvette with distilled or de-
ionized water or other specified solvent. Wipe the cuvette with tissue to remove the
fingerprints and droplets of liquid.
2.2.5 Fit the blank cuvette into the 4-cell holder and place the cuvette in the slot nearest
you. Push the rod so that the cuvette is in the light path. Close the lid.
2.2.6 Set 0.000A or 100%T by pressing 0A/100%T button.
Blanking
Energy low! 546nm
Note: If “Energy low!” is displayed it might indicate the reference is too dark or
the light beam energy from the lamp is too weak.