48 IA_20901569_EN_R9_V1_061223
Maintenance
110. Screw the blind of the gas ballast connec‑
tion tight.
> The mark „GB“ on the blind must point in
direction of the oil reservoir.
111. Insert new insulating material in the gas
ballast connection.
112. Insert the gas ballast cap.
113. Unscrew the oil inlet screw.
114. Fill in oil.
> The oil level must be between the„Min.“ and
„Max.“ markings on the rating plate.
115. Screw in the oil inlet screw.