Questions and Answers

  • A
    amberlesterAug 8, 2025
    What to do if my vacuubrand BVC basic Water Pump fails to start?
    • M
      Melissa ConwayAug 8, 2025
      If your vacuubrand Water Pump fails to start and the system becomes vented, there are several potential solutions: * Ensure the pump is properly plugged in and check the fuse controlling the power outlet. * Verify that the power switch is turned on. * If the motor seems overheated, switch the pump off, allow it to cool down, and identify/eliminate the cause of the failure before restarting. * If a fuse is blown, identify the cause of the failure and replace the fuse (located under the cover). * Adjust the aspiration power to “+“ for more vacuum. * Check for and eliminate any leaks in the system. * Replace the filter.
  • D
    davisgregoryAug 12, 2025
    Why is my vacuubrand BVC basic Water Pump running continuously?
    • D
      Daniel Keller PhDAug 12, 2025
      If your vacuubrand Water Pump is running continuously, you can try the following: * Check the tubing, seals, coupling, and blind cap for any issues. Screw them in tightly or replace them if necessary. * Try reducing the underpressure or aspiration power. If the issue persists after these steps, contacting your local distributor might be necessary.
  • K
    Kara DiazAug 16, 2025
    What to do if there is no aspiration and the pipette drips with my vacuubrand BVC basic Water Pump?
    • K
      Kimberly BarnesAug 16, 2025
      If you're experiencing no aspiration at the VHCpro and the pipette is dripping with your vacuubrand Water Pump (after excluding other causes), consider these solutions: * Replace the adapter with an aspiration tube. * Replace the filter. * Check the seals, coupling, and blind cap, and screw them in or replace them if necessary. * Perform maintenance, replacing the diaphragm and valves if needed. * Set the aspiration power to “+“ for more vacuum.
  • C
    campbellmelissaAug 21, 2025
    Why does my vacuubrand Water Pump switch frequently?
    • G
      garrett53Aug 21, 2025
      If your vacuubrand Water Pump is switching frequently, try these solutions: * Check the tubing, seals, coupling, and blind cap. Ensure they are properly screwed in or replace them if necessary. * Replace the filter. * Perform maintenance, which may include replacing the diaphragm and valves.
  • S
    Scott HoustonAug 25, 2025
    What does it mean when my vacuubrand BVC basic runs continuously and all LEDs are flashing?
    • J
      jesseleAug 26, 2025
      If your vacuubrand Water Pump is running continuously and all the LEDs indicating suction power are flashing, the pressure sensor may be defective. Try reducing the suction power using the “-“ key. If the error persists, contact your local distributor.
  • T
    Theresa ParkerAug 29, 2025
    What to do if the level sensor gives a false alarm on my vacuubrand BVC basic?
    • W
      walkerashleyAug 29, 2025
      If the level sensor gives a false alarm or the level alarm occurs at the wrong time with your vacuubrand Water Pump, consider the following: * If the bottle has been replaced or ambient conditions have changed, adjust the level sensor. * Use an original VACUUBRAND BVC bottle or original spare bottle. * Remove any adhesive foil from the bottle. * Dry the bottle and adjust the level sensor. * Clean the bottle carefully and dry it, then adjust the level sensor.
  • A
    Ashley WellsSep 3, 2025
    What to do if the keys on my vacuubrand BVC basic do not react?
    • E
      Erika HarperSep 3, 2025
      If the keys on your vacuubrand Water Pump do not react, it might be because the touch panel keys were touched for an extended period. Wait 20 seconds and try again.

vacuubrand BVC basic Specifications

General IconGeneral
Number of heads1
Inlet connectionHose nozzle DN 6-10 mm
Permissible ambient temperature10 - 40 °C
Motor protectionIP 20
Voltage230V
Frequency50/60 Hz