page 14 of 27
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 25/11/2013 // 999281
Connectgaspressurebottletothepumpinlet.
+ Keepthevalveofthegaspressurebottleclosed.
Remove hose at the pump inlet.
SwitchofftheBVC.
PositiontheBVCinahoodorinstallanexhaustpipeatthe
outlet of the pump to ensure proper disposal of outgoing
gas.
The applicability of the decontamination method and the selection of the ap-
propriate gas (e. g., formaldehyde, ethylene oxide, ammonia, .....) have to be
ensured by the user.
Decontamination of the pump due to gassing
NOTICE