49
3. Select the number of
calibration points at menu item
[2] Points
. You can perform a 1-
point or 2-point calibration.
For a 2-point calibration, you
need two reference
environments. For example,
LiCl and NaCl salt chambers
provide 11% and 75% relative
humidity references.
Note that when performing a 2-
point RH calibration, the first point
requires a < 50% RH humidity
reference, and the second point
must be > 50% RH.
The difference between the two
references must be at least 30%
RH.
4. Remove the filter from the probe
and place the probe in the first
reference environment (first
calibration point).
Take care not to damage the
sensors. Wait 20 40 minutes for
the reading to stabilize.