. You can perform a 1-
point or 2-point calibration. 
 
For a 2-point calibration, you 
need two reference 
environments. For example, 
LiCl and NaCl salt chambers 
provide 11% and 75% relative 
humidity references. 
 
Note that when performing a 2-
point RH calibration, the first point 
requires a < 50% RH humidity 
reference, and the second point 
must be > 50% RH.  
 
The difference between the two 
references must be at least 30% 
RH. 
 
 
4. Remove the filter from the probe 
and place the probe in the first 
reference environment (first 
calibration point).  
Take care not to damage the 
sensors. Wait 20 – 40 minutes for 
the reading to stabilize.