HMP9 Probe
Figure 8 HMP9 Probe Dimensions
Vaisala HUMICAPâ Humidity and Temperature Probe HMP9 is designed for easy installation
into rapidly changing environments where fast response time, measurement performance, and
chemical tolerance are essential.
The probe head can be mounted through thin metal walls using the included cable gland or
mounting grommet. You can also attach the probe head directly using a zip tie. The probe
head should be attached from the point near the black plastic part.
Temperature measurement range -40 ... +120 °C (-40 ... +248 °F)
Operating temperature range for probe body -40 ... +60 °C (-40 ... +140 °F)
Integrated filter (non-replaceable)
Do not damage the probe head by bending, crushing, or striking it.
Avoid overtightening when installing the probe head through a cable gland.
CAUTION!
14 M211982EN-D