Installing HMP9 Through a Cable Gland
Figure 9 Installing HMP9 Probe Head Through a Cable Gland
1 Black plastic part of the HMP9 probe head
2 Nut for tightening the probe in place
3 Base of the cable gland
4 M10×1.5 threads of the cable gland
• M10×1.5 cable gland (included with HMP9 probe)
• Drill with 8.5 mm bit
• M10×1.5 threading tap
• 13 mm wrench
1. Drill a 8.5 mm diameter hole in the installation location.
2. Use a threading tap to create a M10×1.5 thread in the hole.
3. Install the base of the cable gland in the hole and tighten with a 13 mm wrench.
4. Insert the seal of the cable gland in the base and place the nut of the cable gland over the
probe head.
5. Insert the probe head in the cable gland up to the black plastic part of the probe head.
Leave the black plastic part entirely outside the cable gland. Tighten the cable gland to
finger tightness.
6. Tighten the nut of the cable gland with a 13 mm wrench until the probe head stops
moving. Do not overtighten.