EasyManua.ls Logo

Vaisala HydroMet - Battery Status

Vaisala HydroMet
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 ___________________________________________________________ Troubleshooting
VAISALA______________________________________________________________________ 135
logger output gives the DSI486 version as "A", the communication
module used is the earlier version, DSI486.
Battery Status
You can view battery status information by giving the command
battery without parameters.
Example:
/ > battery
Remaining=100
State=FLOAT_CHARGE
U=6.850195
I=2.685547
Type=PB_BATTERY
Capacity=1.200000
Ext.DC=13.132420
Internal temperature=23.511668
Battery voltage (U) and voltage at the +ExtDC terminal are given as
volts, charging current (I) as milliamperes, and capacity as ampere-
hours. The remaining percentage shows how much energy is left in the
battery.
NOTE
If Program flash checksum returns an error, reload the software. If
reloading the software doesn’t work, contact Vaisala.
NOTE
The battery command applies only to the internal battery of the QML
logger up to QML201B.