5 Setting Up the mniIPS2
5.1 Start / Stop
When power is applied to the miniIPS, it will immediately begin to operate according to the settings already
programmed. The most basic level of Stop / Start control is, therefore, by switching power on and off.
When the instrument is running, the miniIPS2 and 2e may be put into set up mode
at any time by typing the ‘#’ character followed by the ENTER key. The device will
respond with a command prompt ‘>’ and wait the next instruction.
When interrupted an error is often seen. This is generated due to the <CR><LF>
usually associated with the # and should not be interpreted as a problem with the
uvSVX
A “watchdog” function is in operation within the miniIPS2 and 2e
If the unit is interrupted with the ‘#’ character, and no further command is received
for a period of 5 minutes, the sensor will automatically begin sampling data using the
existing settings
5.1.1 Output Last Measured Reading
In order to read the last measured reading the command #015 can be used.
The data is output in the selected units and data telegram format.