miniIPS2 and 2e Operating Manual: MANUAL-383434403-1
5.6 Units
• The default output units for the miniIPS2 and 2e are deciBar (dBar), equivalent to 0.1Bar, or
approximately 1m of seawater
• The miniIPS2 and 2e can also output pressure in a number of different units
• It is also possible to present the data in units of metres or feet of seawater
- calculated using the TEOS 10 Simple Pressure / Depth relationship, which assumes “standard” water
density. A Latitude is required - see Set Latitude
5.6.1 Commands
Set calibrated units as follows, where n = 1 to 6
1 = mBar (m) 2 = dBar (d)
3 = Bar (B) 4 = Metres (M)
5 = Feet (f) 6 = PSI (P)
Example: #018;3
sets the units to Bar
Enable / Disable unit character in the Valeport output string, where n=
0=hide units in valeport string
1=display units in valeport string
Available in the following data string formats:
Valeport NMEA ($PIPS) | Valeport Standard | Valeport NMEA ($PVIPS2) | Impact
SubSea
Read units activation status - 0 = OFF | 1 = ON
All output formats are affected by changing the output units.