miniIPS2 and 2e Operating Manual: MANUAL-383434403-1
5.8 Set Latitude
The Latitude is required for an accurate pressure / depth conversion - the relationship is partially
dependent on local gravity, which varies with distance from the equator.
Positive / negative signing for North / South of the equator is not relevant
# Code
Description
#135;nn.nnn
Sets the local operating latitude in decimal degrees
Example: #135;50.426
Sets latitude to 50.426° (50° 25’ 34”)
#136
Read the Latitude set in the instrument
5.9 Standard Set Up
A customer can request a Standard Set Up script to set their instruments up in a particular manner, output
instrument information, perform variance checks and set to run mode etc..
Please contact Valeport for further information
# Code
Description
#999;n
Set an instrument up as specified by a script
Where n = your script number
#999;0 will set the instrument back to factory default settings