Section 7 Setting up
7-6. Inserting disposable patient circuit:
7-6-1. Open hinged door to expose
the docking station.
7-6-2. Hold disposable patient circuit by its
handle, with the delivery tube downward
as shown.
7-6-3. Slide disposable patient circuit
downward into the docking station until
it stops.
7-6-4. Press down firmly to ensure correct
seating.
NOTE: If the hinged door does not close easily, check that the cartridge is installed
correctly and the disposable water path is fully inserted into the docking station.
CAUTION: Do not remove the disposable patient circuit while the unit is operating.
7-6-5. Close door.