Page 30 3001002 Rev. P
Appendix
Guidance and manufacturer’s declaration - electromagnetic emissions
Emissions test
Compliance
Electromagnetic environment-
guidance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration- electromagnetic immunity
IEC 60601-1-2:2001+A1:2004
EN60601-1-2:2001
AS/NZ3200.1.2:2005
Sub Test Passed Parameters
Electro-Static Discharge
EN 61000-4-2:1995,+A1:1998,
+A2:2001
±6kV Contact discharge
±8kV Air discharge
Radiated RF Susceptibility
EN 61000-4-3:2002
80- 2500MHz @ 3 V/m,
1kHz AM 80% modulation
Electrical Fast Transients
EN 61000-4-4:2004
±5kV AC mains,
Surges
EN 61000-4-5:1995, +A1:2001
±0.5,1kV Line to line
±0.5, 1 2kV Line to protected earth
Line Conducted RF Susceptibility
EN 61000-4-6: 1996, +A1:2001
0.15-80MHz @ 3Vrms,
1kHz AM 80% modulation
Power Frequency Magnetics
EN 61000-4-8: 1993, +A1:2001
3A/m @ 50/60Hz PASSED
Voltage Dips and Dropouts
EN 61000-4-11: 2004
Per Standard
The Precision Flow™ uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby equipment.
The Precision Flow™ is intended for use in the electromagnetic environment specified below. The customer
or the user of the Precision Flow™ should assure that it is used in such an environment.
The Precision Flow™ is suitable for use in all establishments other
than domestic, and may be used in domestic establishments and
those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes,
provided the following warning is heeded:
Warning: This equipment/system is intended for use by healthcare
professionals only. This equipment/system may cause radio
interference or may disrupt the operation of nearby equipment. It
may be necessary to take mitigating measures, such as
re-orienting or relocating the Precision Flow™ or shielding the
location.