14
The user must provide a stable power supply for the product. In case of severe fluctuations in the power
supply, a power regulator or UPS power supply should be used for adjustment purpose.
Warning:
All the system components shall be powered off before connecting them. Only appropriate cables can be
used for the product; or else, the product may be damaged or malfunction.
The product can only be connected to a medical display screen that has been qualified as per IEC 60601-1.
The product can be operated only under the conditions that meets the requirements given in "Product
Specifications and Recommended Accessories", "Operating Environment", "Storage and Transportation
Environment" sections; or else, malfunction, impaired safety and/or equipment damage may be caused.
Do not bend, pull, twist or squeeze the cables sharply.
Do not place any equipment on top of the product; or else, the product may be damaged.
Do not install the product near any strong electromagnetic radiation source (e.g. microwave therapy
equipment, shortwave therapy equipment,and MRI); or else, the product may malfunction.
Make sure the product completely dry before connecting the endoscope connector to it; or else, electric
shock or equipment damage may be caused.
When using the product for the first time, please refer to this User's Manual and install the product
according to the following steps. See Chapter II for the positions corresponding to the numbers referenced
below:
3.1.1 Open the product packaging to ensure that all the components are provided. See Chapter I for the
attached components.
3.1.2 Carefully check the product and all the components for damage (and without wear). Never the product
in case of any damage.
3.1.3 Open the bracket (12) of the product and place it on a solid surface.
3.1.4 Connect the power supply to the power adapter and the power jack (8).
3.1.5 Insert the endoscope connector into the endoscope socket (11) properly.
3.1.6 Connect the power cable to the AC power socket.
3.1.7 Short press the power button (1) to turn on the product.
3.1.8 Set the product with reference to Section 3.3.4.
Note: Some key functions need to enter the user password when using,The initial user password of this
product is 123456, you can reset the user password according to the instructions in the manual.