Appendix 1. Electromagnetic Compatibility
The following cable information is given for reference only.
Max. cable length
Shielded/unshielded
Important information on electromagnetic compatibility (EMC)
The product shall be subjected to special precautions regarding EMC, and the EMC information provided in its
user's manual shall apply when using it. The product complies with IEC 60601-1-2:2014 in terms of immunity and
emission. However, the following special considerations shall be observed:
The product with the following basic performance is suitable for the professional medical environment, except for
any places that is near the RF shielding room of the ME system for magnetic resonance imaging (MRI) with high
electromagnetic interference intensity.
Basic performance
The product is designed to provide images for observation, excluding short-term automatically recoverable
reduction caused by electromagnetic interference.
Warning: It is required to avoid using the product next to or stacked with other equipment; or else, improper
operation may be caused. If you have to do so, the product and other equipment should be observed carefully to
verify that they can operate properly.
The use of accessories, transducers or cables other than those specified or provided by the manufacturer of the
product may increase electromagnetic radiation or reduce electromagnetic interference, and cause improper
operation.
Warning: When using any portable radio frequency communication equipment (including antenna cables, external
antennas and other peripheral equipment), such equipment should not be placed over 30 cm (12 inches) away
from any part of the product, including the cables specified by the manufacturer; or else, the performance of such
equipment may be impaired.
Statements: The product is designed compatible with the high-frequency surgical equipment, i.e. it is allowed to
keep working or standby near the high-frequency surgical equipment.
In case of interrupted AC input voltage, the product will shut down, and once the power supply is restored, it can
be manually restored by the operator. The degradation in such case is acceptable because it will neither cause
unacceptable risks nor impair basic safety or basic performance.
The degradation due to electrostatic discharge or electrical fast transient/burst is acceptable because it will neither
cause unacceptable risks nor impair basic safety or basic performance.
The vertical-line flashing on the screen can automatically restore to the previous state.