EMI compliance (Table 1)
Table 1. Emission
Electromagnetic Environment
Professional medical environment
Professional medical environment
Voltage fluctuation
and Flicker
Professional medical environment
Caution: The emission characteristics of the product make it suitable for use in industrial areas and hospitals
(CISPR 11 Class A). If being used in the residential environment (CISPR 11 Class B required usually),
the product may not provide adequate protection for radio frequency communication services. The
user may be required to take mitigation measures, for example, relocating or redirecting the
product.
EMS compliance (Tables 2–5)
Table 2. Chassis ports
Professional medical environment
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated,
radio-frequency,
electromagnetic field
3 V/m
80 MHz–2.7 GHz
80% AM at 1 kHz
Fields near the
radio-frequency
wireless
communication
devices
Rated power frequency
magnetic field
Table 3. Fields near the radio-frequency wireless communication devices
Professional medical environment
Pulse modulation 18 Hz, 27 V/m
FM, ±5 kHz deviation, 1 kHz sine, 28 V/m
Pulse modulation 217 Hz, 9 V/m