23
7. Repeat Steps 1 to 6 until the Digital Video Monitor is clean.
7.2. Disinfection
1. Disinfection Solution: isopropanol (alcohol) concentration 7080%;
2. Preparation: add 80 cc of 95% isopropanol (alcohol) to 20 cc of purified water (PURW) (or use the EPA
registered medical disinfectant wet wipes containing at least 70% isopropanol; all the appropriate safety
precautions and manufacturer’s instructions must be followed).
3. Use a piece of sterile gauze dipped in the alcohol mixture described above to wipe the surface of the
product for about 15 minutes (approximately once every 2 minutes). The isopropanol should be handled
according to appropriate safety procedures. The gauze should be moist without dripping because any liquid
may impair the electronics inside the product. Always pay close attention to the buttons, housing, slots and
gaps on the product. The sterile cotton swabs should be used for the disinfection in these areas.
8. Maintenance and Disposal
8.1. Storage
After being cleaned and disinfected, the product must be subjected to the pre-inspection procedures
described in Chapter 5. The product must be stored during the period between two operations according to
local guidelines and the storage environment requirements in this manual.
8.2. Battery Maintenance
In order to extend the service life of the battery, it is recommended to fully charge the product at least once
every three months. The battery can last over 6 hours. The temperature range during charging should be
controlled within 1035 °C.
If the battery needs to be replaced, never refit any accessory of the product at will, and instead, contact your
local dealer or Vathin Medical.
8.3. Return the Digital Video Monitor for repair
Vathin Medical shall have the right to request the technical department or equivalent at the customer's site to
repair the product under the proper guidance of Vathin Medical. Any defective Vathin Digital Video Monitor
must be disposed of by the person authorized by Vathin Medical. Circuit diagrams, component part lists,
descriptions, calibration instructions, or other information are available on request by auhorized SERVICE
PERSONNEL.
In order to prevent infection, it is strictly forbidden to transport contaminated medical equipment. The
medical equipment must be disinfected on site before being shipped to Vathin medical. The cleaning and
disinfection procedures described in Chapter 7 must be followed. Vathin Medical reserves the right to return
contaminated medical equipment to the sender.