62
9 Diagnosis, asset management and service
VEGADIF 85 • Modbus and Levelmaster protocol
53571-EN-230814
Modicationofaparameter
Switch-onandswitch-otimes
Status messages (according to NE 107)
Error messages (according to NE 107)
ThedataarereadoutviaaPCwithPACTware/DTMorthecontrol
system with EDD.
9.3 Asset Management function
The instrument features self-monitoring and diagnostics according
toNE107andVDI/VDE2650.Inadditiontothestatusmessagesin
the following tables there are more detailed error messages available
under the menu item " Diagnostics" via the respective adjustment
module.
The status messages are divided into the following categories:
Failure
Function check
Outofspecication
Maintenance required
and explained by pictographs:
41 2 3
Fig. 45: Pictographs of the status messages
1 Failure - red
2 Out of specication - yellow
3 Function check - orange
4 Maintenance required - blue
Malfunction (Failure):
Due to a malfunction in the instrument, a fault signal is output.
This status message is always active. It cannot be deactivated by the
user.
Function check:
The instrument is being worked on, the measured value is temporarily
invalid (for example during simulation).
This status message is inactive by default.
Outofspecication:
Themeasuredvalueisunreliablebecauseaninstrumentspecication
was exceeded (e.g. electronics temperature).
This status message is inactive by default.
Maintenance required:
Duetoexternalinuences,theinstrumentfunctionislimited.The
measurementisaected,butthemeasuredvalueisstillvalid.Planin
Status messages