6.6 Saving the parameter adjustment data
It is recommended noting the adjusted data, e.g. in this operating
instructions manual and archive them afterwards. They a re hence
available for multiple use or service purposes.
If VEGAFLEX 62 is equipped with an indicating and adjustment
module, the most important data can be read out of the sensor into
indicating and adjustment module. The procedure is described in the
operating instructions manual "Indicating and adjustment module" in
the menu item "Copy sensor data". The data remain there permanently
even if the sensor power supply fails.
If it is necessary to exchange the sensor, the indicating and adjustment
module is inserted into the replacement instrument and the data are
written into the sensor under the menu item "Copy sensor data".
40 VEGAFLEX 62 4 20 mA/HART two-wire
6 Set up with the indicating and adjustment module PLICSCOM
31837-EN-090831

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vega VEGAFLEX 62 and is the answer not in the manual?

Vega VEGAFLEX 62 Specifications

General IconGeneral
Accessory TypeSensor
ApplicationLevel measurement of liquids
Measuring principleGuided wave radar
Measuring rangeUp to 75 m (246 ft)
Process temperature-40 to +200 °C
Process pressureUp to 40 bar
Accuracy±2 mm
Output signal4 ... 20 mA/HART
Power supply12 ... 36 V DC
Protection classIP66
MaterialStainless steel 316L, Alloy C22
Connectionflange