Run another cycle, if issues
persist, call maintenance
Check pump for correct operation (use
engineers function). Check in-line fuse to
pump.
Check pipework for blockages
Check drain valve is sealing correctly
Check if motor turns. Check that safety
switch is operating correctly. Check for
error codes on inverter
Check foot sensor operation, recalibrate
if required
The lid can be manually opened by
removing the front cover and manually
withdrawing the lid lock using the lever
beneath the lock. The lid can then be
opened against the magnetic clutch.
Open the small covers next to the hinges
to check operation of the lock, motor,
clutch and to inspect the connection
between motor and lid