Chapter A Appendix A: RCI commands
AC CAP Measure
22166183 R001 PSD90-3 Universal AC/DC Fuel Quantity Test Set Operation Manual FINAL
March 2023 Subject to Export Control; see Cover Page for details. A - 53
A.6.2 Read measurement
Co mmand :ACCAP:RANGE?
De s cript ion Command will return the current r ange in ρF.
Ex a mple :ACCAP:RANGE?
Re spons e Ex a mpl e 40000
Co mmand :ACCAP:AUTO <is_auto>
De s cript ion Command sets the current ranging mode for the ACCAP
measurement.
Para m ete rs <is_auto>
– 1 for AUTO
– 0 for Manual
Ex a mple :ACCAP:AUTO 1
Co mmand :ACCAP:AUTO?
De s cript ion Command r eturns the current ranging mode:
– 1 for AUTO
– 0 for Manual
Ex a mple :ACCAP:AUTO?
Re spons e Ex a mpl e 1
Co mmand :ACCAP:READ:CAP?
De s cript ion Command r eturns the most recent reading, in ρF.
Re spons e Ex a mpl e 25.251
Co mmand :ACCAP:READ:PHASE?
De s cript ion Command r eturns the most recent phase, in degrees..
Re spons e Ex a mpl e 1